dc.description.abstract |
The purpose of this research was to develop a clinically validated portable NPWT device with low power consumption
features. A new NPWT device having dimension of 30 cm x20 em <licm (Length x Width x Hight) was clinically tested to
evaluate its performance. The device is capable of working continuously for more than 24 hours under pressure range
from -50 mmHg to -250 mmHg with ± 3 mmHg accuracy. After several clinical tests, it could be confirmed that the
wound healing rate has been increased significantly with the aid of developed NPWT device. |
en_US |
dc.language.iso |
en |
en_US |
dc.title |
Design and testing a portable negative pressure wound therapy (NPWT) device |
en_US |
dc.type |
Conference-Full-text |
en_US |
dc.identifier.faculty |
Engineering |
en_US |
dc.identifier.year |
2013 |
en_US |
dc.identifier.conference |
National engineering conference engineering research for national building |
en_US |
dc.identifier.place |
University of Moratuwaw, Moratuwa, Sri Lanka |
en_US |
dc.identifier.pgnos |
128-131 |
en_US |
dc.identifier.proceeding |
Proceeding of the 18th annual research symposium |
en_US |
dc.identifier.email |
pdwdgama@gmail.com |
en_US |
dc.identifier.email |
hans.gray@gmail.com |
en_US |
dc.identifier.email |
ranamajp@gmail.com |
en_US |
dc.identifier.email |
thusitha@mech.mrt.ac.lk |
en_US |
dc.identifier.email |
mandika59@hotmail.com |
en_US |