Design and testing a portable negative pressure wound therapy (NPWT) device

dc.contributor.authorWelgama, WPD
dc.contributor.authorGray, HA
dc.contributor.authorAmarasinghe, YWR
dc.contributor.authorWijeyaratne, SM
dc.contributor.authorSugathapala, AGT
dc.date.accessioned2014-03-21T14:16:17Z
dc.date.available2014-03-21T14:16:17Z
dc.date.issued2013
dc.description.abstractThe purpose of this research was to develop a clinically validated portable NPWT device with low power consumption features. A new NPWT device having dimension of 30 cm x20 em <licm (Length x Width x Hight) was clinically tested to evaluate its performance. The device is capable of working continuously for more than 24 hours under pressure range from -50 mmHg to -250 mmHg with ± 3 mmHg accuracy. After several clinical tests, it could be confirmed that the wound healing rate has been increased significantly with the aid of developed NPWT device.en_US
dc.identifier.conferenceNational engineering conference engineering research for national buildingen_US
dc.identifier.emailpdwdgama@gmail.comen_US
dc.identifier.emailhans.gray@gmail.comen_US
dc.identifier.emailranamajp@gmail.comen_US
dc.identifier.emailthusitha@mech.mrt.ac.lken_US
dc.identifier.emailmandika59@hotmail.comen_US
dc.identifier.facultyEngineeringen_US
dc.identifier.pgnos128-131en_US
dc.identifier.placeUniversity of Moratuwaw, Moratuwa, Sri Lankaen_US
dc.identifier.proceedingProceeding of the 18th annual research symposiumen_US
dc.identifier.urihttp://dl.lib.mrt.ac.lk/handle/123/9849
dc.identifier.year2013en_US
dc.language.isoenen_US
dc.titleDesign and testing a portable negative pressure wound therapy (NPWT) deviceen_US